Analytical Gas Systems Catalogue

9 www.parker.com/gsfe High Purity Nitrogen Generators For analytical instruments including ICP Weights and Dimensions Technical Data Model Height (H) Width (W) Depth (D) Weight (with compressor) Weight (without compressor) mm in mm in mm in kg lb kg lb HPN2 range 869 34.2 345 13.6 667 26.3 95 189.6 86 209.4 Product Selection Model Flow Rate Purity* Inlet air @ 8 to 9.9 bar (116 to 143 psi) ** Oulet Pressure Integral Compressor L/min % L/min bar g psi g HPN2-5000 5 >99.999 110 5 72.5 No HPN2-5000C 5 >99.999 n/a 5 72.5 Yes HPN2-12500 12.5 >99.5% 129 5 72.5 No HPN2-12500C 12.5 >99.5% n/a 5 72.5 Yes *Purity with respect to oxygen ** Model HPN2-12500 min inlet 9 bar (130 psi) Note: Add suffix ‘E’ for 207-253V 50/60Hz ie. HPN2-7000-E Add suffix ‘W’ for 103 -126V 60Hz ie. HPN2-7000-W * Non compressor models only Ambient Temperature Range 15 - 25°C (55 - 77°F) Inlet Air Quality * Clean dry compressed air ISO 8573-1:2001 Class 2.-.1 Supply Voltage Range 103 - 126V 60Hz / 207 - 253V 50/60Hz Port Connections Outlet Inlet * ¼" Compression Fitting ¼" Compression Fitting Preventative Maintenance Preventative Maintenance Kit - all models Part Number Change Frequency Filter kit - non compressor models 606272557 12 Months Filter kits - compressor models 606272559 12 Months Compressor Kit 230V models 606272581 8,000 hours or 24 months (which ever comes first) Compressor Kit 120V models 606272583 8,000 hours or 24 months (which ever comes first) Optional Extras Description Part Number Required for Installation kit IK7694 Suitable for all HP nitrogen generators

RkJQdWJsaXNoZXIy MTQ4NTgyMQ==